Procedure:
Blood Glucose
Monitoring
Product: Roche Diagnostics AccuData GTS
(formerly Boehringer Mannheim)
State: Pennsylvania
Accrediting organization: CAP
NCCLS format: Yes |
Point-of-Care
Testing
<Very
Good Health Hospital>
Blood
Glucose Monitoring
PRINCIPLE:
Point of Care
Testing (POCT) refers to those analytical patient testing
activities provided within the hospital, but performed outside the
laboratory. Activities of the POCT program must comply with all
current standards for laboratory accreditation, regardless of the
scope of testing. Designated laboratory personnel have centralized
the coordination of the POCT program.
The Boehringer
Mannheim Corporation (BMC) Accudata Glucose Testing Station (GTS)
using the Accu-check Advantage monitor is used for blood glucose
determinations at the bedside. This system uses a drop of whole
blood placed on a reagent strip and inserted into a monitor to
obtain the glucose value. The monitor’s test reaction is based
on the enzyme glucose dehydrogenase converting the glucose in the
blood sample to gluconolactone. This reaction liberates an
electron that reacts with a coenzyme electron acceptor, the
oxidized form of the mediator hexacyanoferrate (III), forming the
reduced form of the mediator, hexocyanoferrate (II). The Advantage
test strip employs the electrochemical principle of biamperometry.
The monitor applies a voltage between two identical electrodes,
which causes the reduced mediator formed during the incubation
period to be reconverted to an oxidized mediator. This generates a
small current that is read by the monitor.
Operation of
the Accu-Chek Advantage monitor can be performed by the following
trained individuals: Registered Nurse (RN), Graduate Nurse (GN),
Licensed Practical Nurse (LPN), Clinical Technical Associate (CTA),
Patient Support Associate (PSA), and a Clinical Technician (CT).
These individuals can only perform a blood glucose that has been
initiated with a physician’s order. These individuals must be
trained by a BMC representative or a <Very Good Health
Hospital> trainer initially and must show competency at
least once per year.
Back to
top
SPECIMEN
COLLECTION AND HANDLING:
Personal
Protective Equipment (PPE) such as gloves, gown and/or goggles
must be worn when handling patient blood samples and quality
control samples. Please refer to the Hospital Administrative
Policy xxx- OSHA Standard Compliance for more specific
information.
I.
Specimen:
Samples
obtained for use on the Accu-Chek Advantage monitor must be
capillary specimens unless otherwise stated by the laboratory.
A minimum sample size of 9uL and a maximum sample size of 50
uL is required.
II. Special
Considerations:
*If
possible the patient should wash his or her hands prior to
drawing a capillary sample from the fingertip. If this is not
possible, make sure the finger is cleaned with an alcohol swab
before performing the fingerstick.
*The
capillary sample must be tested immediately after collection.
REAGENTS
AND EQUIPMENT:
- Accudata
GTS with Accu-Chek Advantage Monitor
- Advantage
comfort curve test strips (Cat. No. 2030365)
- Advantage
controls (Cat. No. 2030390)
- Safe-T-Pro
disposable lancing devices (Cat. No. 951) or pediatric lancing
devices.
- Alcohol
swab
- Cotton
ball, tissue, or gauze for wiping finger after stick
- Disposable
latex gloves
Back to
top
STORAGE
AND STABILITY:
Advantage
monitor:
The monitor
should be stored between 14oC and 40oC (57oF
to 104oF) away from moisture.
Advantage test
strips:
The Advantage
test strips must be stored between 14oC and 40oC
(57oF to 104oF). They must be stored in the
same tightly capped vial in which they are packaged. The vial cap
must immediately be replaced after the removal of a test strip.
The test strips are stable until the expiration date printed on
the vial. The date the test strips are received and opened must be
written on the vial label. Test strips should be kept at less than
85 % humidity for the most accurate results.
Advantage low
and high controls:
The Advantage
low and high control solution vials must be stored at room
temperature. The control solutions are stable until the expiration
date printed on the vial or 3 months after opening, whichever
comes first. The date the control solutions are received and
opened must be written on the vial label along with the new
expiration date if applicable.
CLEANING
AND MAINTENANCE:
- Cleaning of
the Accu-chek GTS and the Advantage monitor is recommended
whenever blood or body fluids have spilled on the monitor.
Either a 10% fresh bleach solution or 70% isopropyl alcohol
can be used to gently wipe the areas that are contaminated.
Operators should keep moisture away from the test strip guide
and only use a soft cloth on the monitor display areas.
Comment code 03 (which translates to "Clean
Monitor") must be used to document cleaning of the
monitor. This information will be reviewed by the Medical
Director on a monthly basis to track problems and/or concerns
with cleaning the monitor.
- Six size C
batteries for the Accu-chek GTS can be obtained from the
storeroom. The monitor will display "change
batteries" when necessary. The Advantage monitor requires
2-3 volt batteries obtained from the storeroom. The Advantage
monitor will display a small battery icon when necessary. The
comment code 06 (which translates to "Change
batteries") must be used to document that batteries were
changed. This information will be reviewed by the Medical
Director on a monthly basis to track problems and/or concerns
with changing batteries.
Back to
top
CALIBRATION:
The Accu-chek
Advantage monitor must be calibrated each time a new lot number of
Advantage test strips is opened. Otherwise, calibration should be
performed at least every six months or more often if deemed
necessary.
Each box of
test strips has a code key that contains the information needed to
calibrate the instrument. This code key is to be left in the GTS
unit until calibration is performed in six months or the monitor
has prompted an operator to calibrate.
Calibration
verification using the linearity test kit (Cat. No. 987) must be
performed when reagent lots are changed, when indicated by quality
control data, after major maintenance or at least every six
months. The reportable range of 60-400 mg/dL for adults and 40-400
mg/dL for infants must be verified from the linear testing
procedure and approved by the Medical Director.
Back to
top
QUALITY
CONTROL:
Quality
Control vials (one low and one high solution) must be run on each
Advantage monitor every 24 hours that the monitor is in use. The
Accudata GTS has a 24 hour lock-out feature that requires both
controls to be run producing values that fall within the
acceptable control range before any other tests can be performed
on the monitor. If the quality control values are within the
acceptable range printed on the test strip vial, the user can
proceed to patient testing. If the quality control values are out
of the acceptable range printed on the test strip vial, the user
must do the following:
- Check the
expiration date to verify that the control vial is stable. If
it has expired, dispose of the control vial in the proper
hazardous waste receptacle and obtain a new vial. Enter the 2
digit corrective action comment code into the GTS and run the
new vial of control. See the procedural notes with detailed
information on the proper use of comment codes.
- If the
control is still out of acceptable range, enter the proper 2
digit corrective action code again. Mix the control by gently
inverting the vial 4 to 6 times and repeat.
- If the
control is still out of acceptable range, refer to the
troubleshooting section of the User’s Manual or call the BMC
Medical Service Center at 1-800-440-3638.
The quality
control values are stored in the monitor until the information is
downloaded weekly by the POCT Coordinator. The data obtained
through this process will include: test result, operator ID,
control level, control lot information, test strip code
information, test time and date, and comment codes.
Back to
top
PROCEDURE:
Calibration:
- Remove the
code key from the Advantage test strip box.
- Compare the
3-digit number on the code key with the number on the test
strip vial.
- Make sure
the Accudata GTS is turned OFF.
- Remove the
old code key form the Accu-chek Advantage module and snap the
new code key (slots facing towards the monitor) into the code
key slot.
- Leave the
code key in the Accudata GTS base and turn the monitor on by
pressing the blue power on button.
- Verify that
the code number on the monitor display corresponds to the code
number on the vial of test strips used.
- If one
cannot verify that the code number on the monitor and the code
number on the test strip vial correspond, the monitor must be
recoded (recalibrated).
- Recalibrate
the Accudata GTS by opening a new vial of test strips and
placing the code key into the monitor once it has been turned
off.
- If
recalibrating is unsuccessful do not use the monitor and call
the Medical Service Center at 1-800-440-3638.
Back to
top
Calibration
Verification using Linearity Kit:
- Press the
blue power ON button located on the Accudata GTS base.
- Once the
date, time and code number (test strip lot number) are
displayed on the LCD panel located above the number key pad,
the instrument is ready to begin.
- Select the MENU
choice on the right by pressing that button.
- Select QUALITY
CONTROL by pressing the arrow button located below this
selection.
- Select LINEAR
next and verify that the test strip vial corresponds to the
code number on the GTS.
- Select YES
if the code corresponds and proceed. If the code does not
correspond, follow the calibration procedure above.
- Enter
operator id. and press ENTER.
- Verify the
linear lot displayed on the GTS. Select yes if the lot number
is the number you wish to use. Select NO and enter the correct
lot number if the lot number displayed is not the lot number
you wish to use.
- Enter 5 or
6 for the number of solutions to be used from the linearity
kit.
- Perform the
linearity testing by following steps 9-15 below in the Quality
Control procedure.
- Select YES
to perform another linearity test until all levels are
complete. Select NO when linearity testing is complete
and the GTS will turn off.
Back to
top
Quality
Control:
- Press the
blue power ON button located on the Accudata GTS base.
- Once the
date, time and code number (test strip lot number) are displayed
on the LCD panel located above the number key pad, the
instrument is ready to begin.
- If controls
have already been run, the monitor will display the CTRLS,
REVIEW, and PAT options. If the controls have not been run, the
monitor will display RUN CONTROLS with the CTRLS and REVIEW
options.
- Select the CTRLS
choice on the left by pressing the arrow button located below
this selection.
- The
instrument will ask you to verify that the code number on the
test strip vial corresponds to the code number on the GTS
display. Select YES by pressing the arrow button located
below this selection, if these two numbers correspond. Select NO
by pressing the arrow button located below this selection,
if the code number displayed does not correspond to the code
number on the vial of strips you have chosen to use. See the
Procedure for entering test strip codes located in the Reagents
section of the User’s Manual.
- Enter the
operator ID by pressing the correct sequence of number keys,
then press enter.
- Select the
desired level of control by pushing the LEVEL1 button or the
LEVEL2 button.
- Verify that
the lot number of control solution on the vial corresponds with
lot number displayed on the GTS and select YES by
pressing the arrow button located below this selection, if this
number is correct. Select NO by pressing the arrow button
located below this selection, if the lot number does not
correspond. See the procedure for entering correct control lot
numbers located in the Quality Control section of the User’s
Manual.
- Remove a test
strip from the vial and immediately replace the cap.
- When the
flashing strip icon appears on the monitor display, gently
insert the test strip (silver bar end) into the bottom end of
the monitor. The yellow target area must be facing up.
- When the
flashing drop icon appears on the monitor display, apply one
drop of control solution to the center of the yellow target
area. Note: Do not allow the tip of the control solution to
touch the test strip.
- Once the
control solution has been added to the test strip, a small box
will flash on the monitor display and in 40 seconds the blood
glucose value will appear.
- Enter a
comment code if necessary, then press ENTER. Comment
codes provide evidence of corrective action when control results
exceed defined tolerance limits. See procedural notes for a
complete list of comment codes.
- Remove the
test strip from the monitor and dispose of the strip and the
lancet in the proper biohazard receptacle.
- Select YES
on the GTS display to perform the other control test level.
Verify the operator ID number and go to step 6 of this
procedure. If no other control tests need to be performed,
select NO on the GTS display and the instrument will
automatically turn off.
- The control
results are stored in the GTS until the information is
downloaded weekly by the POCT Coordinator. This information can
also be retrieved by using the review function on the Acudata
GTS.
Back to
top
Patient
Testing:
- Press the
blue power ON button located on the Accudata GTS base.
- Once the
date, time and code number (test strip lot number) are displayed
on the LCD panel located above the number key pad, the
instrument is ready to begin.
- Select the PAT
choice on the right by pressing the arrow button located below
this selection.
- The
instrument will ask you to verify that the code number on the
test strip vial corresponds to the code number on the GTS
display. Select YES by pressing the arrow button located
below this selection, if these two numbers correspond. Select NO
by pressing the arrow button located below this selection,
if the code number displayed does not correspond to the code
number on the vial of strips you have chosen to use. See the
procedure for entering test strip codes located in the Reagents
section of the User’s Manual.
- Enter the
operator ID by pressing the correct sequence of number keys,
then press enter.
- Enter the
patient ID (9 digit medical record number) by pressing the
correct sequence of number keys, then press enter. If the wrong
patient ID number is entered, comment code number 08 shall be
added to this result.
- Remove a test
strip from the vial and immediately replace the cap.
- When the
flashing strip icon appears on the monitor display, gently
insert the test strip (silver bars end) into the bottom end of
the monitor. The yellow target area must be facing up.
- When the
flashing drop icon appears on the monitor display, obtain a
blood sample with the Safe-T-Pro lancing device or pediatric
lancing device. See the Laboratory fingerstick/heelstick policy
as well as Infection Control Practice #12 for Phlebotomy
Procedures for more details. Briefly touch the drop of blood to
the center of the yellow target area. Completely cover the
yellow target area with blood. Do not smear the blood on the
yellow target area.
NOTE:
The monitor may be removed from the GTS to make it easier to
insert the test strip or to dose the test strip. The
Advantage monitor should not be removed from the GTS before
the flashing strip icon appears. Once the testing is
completed, the monitor must be placed back on the GTS before
the next test is performed.
- Once the
blood has been added to the test strip, a small box will flash
on the monitor display and in 40 seconds the blood glucose
value will appear. The value on the monitor display should
correspond with the value on the GTS display. Verify this
result on both displays.
- Enter a
comment code if necessary, then press ENTER. Comment codes
provide evidence of corrective action when patient results are
questioned and/or need to be repeated. See procedural notes
for a complete list of comment codes.
- Remove the
test strip from the monitor and dispose of the strip and the
lancet in the proper biohazard receptacle.
- Select YES
on the GTS display to perform another test. Verify the
operator ID number and go to step 6 of this procedure. Select NO
on the GTS display to turn the instrument off.
- Once a
patient result is displayed on the monitor, it must be
recorded on the Daily Clinical Flow Sheet. The following
information can be obtained from each patient by using the
review function on the Accudata GTS: test result, operator ID,
patient ID, test strip code information, test time and date,
and comment code.
Back to
top
PROCEDURAL
NOTES:
*The comment
codes available to the users after each patient test and each
control run include:
- No action needed
- Repeat test
- Wrong control run
- Clean monitor
- Replace test strips
- Replace control solution
- Change batteries
- Operator error
- Wrong patient ID
- Wrong operator ID
- Monitor taken out of service
- Patient fasting
- Patient not fasting
- Physician notified, lab blood
glucose drawn
A comment code
must be entered or the test will automatically default to 00 (No
action needed). Comment codes provide corrective action
information.
*Contact
Isolation Patients: If a patient with contact isolation requires
blood glucose monitoring at the bedside, an Accudata GTS with
Accu-Chek Advantage monitor should be designated for this patient
and kept in the isolation room. Extra monitors will be available
in the laboratory at both sites for use on these patients. These
monitors must not leave the room of the isolation patient until
the patient’s discharge. Proper cleaning and disinfecting the
monitor by unit personnel is required before sending it back to
the laboratory. See Contact Isolation Patients Protocol for
detailed information.
EXPECTED
PATIENT RESULTS:
The results of
patient testing will be directly recorded on the Daily Clinical
Patient Flow Sheet. The precision of the monitor is expected to be
+ 10%. If this precision after a repeat is not seen on the
monitor and the operator believes this could be an unusual or
unexpected result, the physician must be notified to determine if
he wants a lab drawn blood glucose performed.
The Accudata
GTS with Accu-Chek Advantage Monitor measures blood glucose levels
from 10 to 600 mg/dL. A reading of LO on the monitor indicates a
result below 10mg/dL and a reading of HI on the monitor indicates
a result above 600 mg/dL. The normal fasting adult blood glucose
range for a non-diabetic is 70-105 mg/dL.
Any result
< 60mg/dL or > 400 mg/dL must be repeated. If the result is
still <60mg/dL then:
The
Hypoclycemia Protocol must be followed at <Hospital 1>,
<Hospital 2>, and <Hospital 3>
Hospitals. If the result is still >400mg/dL the result must be
reported to the physician and a laboratory blood glucose must be
drawn.
For infants
located in the NICU and Newborn Nursery, a result of <40mg/dL
must be reported to the physician and glucose feeds according to
the Standard of Care Protocol must be initiated. The specific
comment code for doing this must be entered into the monitor when
indicated to document that the proper protocol was followed.
Before new
lots of reagents are ready for use by operators, parallel studies
must be performed. These studies must compare the previous lot
number to the new lot number so that similar quality control and
patient reference ranges are used throughout <Very Good
Heath Hospital>. Data from at least 10 different floors to
include at least 100 results must be collected by the POCT
Coordinator to perform this comparison study. If the new lot of
reagents does not correlate well according to the Medical
Director’s analysis, the monitors must be re-calibrated and the
comparison study should be performed again. If the new lot still
does not correlate well, a new lot of reagents should be obtained
from the Boehringer-Mannheim Corporation.
LIMITATIONS
AND INTERFERING SUBSTANCES:
- Use only
Advantage test strips (Cat. no. 787) for testing capillary
blood on the Accu-chek Advantage Blood Glucose monitor.
- Patient’s
hematocrit levels must be between 20 and 65% for this monitor
to provide an accurate result. Blood glucose levels above
200mg/dL require hematocrits to be 25 to 55% to provide an
accurate result. If hematocrits do not fall within these
ranges, a lab blood glucose must be drawn in place of the
bedside glucose.
- If any
yellow mesh is still visible on the test strip after the drop
of blood has been applied, more blood may be applied within 15
seconds of the first dose. If more than 15 seconds have
passed, the test result may be erroneous and the test must be
repeated.
- Galactose
in excess of 10 mg/dL may give falsely elevated results versus
a whole blood reference. No interferences were found with
other reducing sugars.
- Uric acid
concentrations greater than 10mg/dL in hypoglycemic patients
(blood sugars <70 mg/dL), and uric acid > 16 mg/dL in
euglycemic and hyperglycemic patients (blood sugars >70 mg/dL)
may give falsely elevated results versus a whole blood
reference.
- In
situations of decreased blood flow, fingerstick blood testing
may not be appropriate because it may not reflect the
patient’s true physiological state. Examples would include
but are not limited to: severe dehydration caused by diabetic
ketoacidosis or the hyperglycemic hyperosmolar nonketotic
state, hypotention, shock, or peripheral vascular disease.
Back to
top
REFERENCES:
Accudata GTS
User’s Manual, Boehringer Mannheim Corporation. 1995.
Total Quality
Management Policy Manual, Boehringer Mannheim Corporation. 1997.
NCCLS GP2-A3
Guidelines for Clinical Laboratory Technical Procedure Manuals
Fundamentals
of Clinical Chemistry, third edition, Teitz, 1987
Prepared
January 1998 |
. |
Medical
Director |
Date |
Approved |
. |
. |
Reviewed |
. |
. |
Reviewed |
. |
. |
Reviewed |
. |
. |
<%
'at this time we will put all the messages entered in
'the table
While Not rs.EOF
%>
Posted
by
Company/Hospital <%=rs("strCompany")%>
Entry time <%=rs("dtmPostedTDS")%>
|
|