Procedure |
Activated
Clotting Time of Whole Blood on the Hemochron Models 401 and 801 |
Prepared by |
Date Adopted |
Supersedes Procedure |
|
|
|
Review Date |
Revision Date |
Signature |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Distributed to |
No.
of Copies |
Distributed to |
No.
of Copies |
Dialysis
Department |
1 |
Pathology
Department |
1 |
Pediatric
Intensive Care |
1 |
Cardiovascular
Surgery |
1 |
ECMO
Department |
1 |
|
|
Cardiac
Cath Lab |
1 |
|
|
PRECAUTION:
WHILE PERFORMING THIS PROCEDURE, THE FOLLOWING SAFETY MEASURES
MUST BE TAKEN: GLOVES MUST BE WORN; SMOKING, EATING, DRINKING, APPLICATION OF
COSMETICS, AND MANIPULATION OF CONTACT LENSES ARE PROHIBITED IN ALL TECHNICAL
WORK AREAS. REFER TO THE UNIVERSAL
PRECAUTIONS POLICIES. |
Complexity:
Moderate
I. Method:
The
activated clotting time (ACT) is measured with the Hemochron Model 401 or 801
instruments. The ACT test is commonly used for heparin anticoagulation
monitoring during bypass surgery, neonatal Extracorporeal Membrane Oxygenation
(ECMO), hemodialysis, radio frequency ablation, and critical care.
II. Principle:
Heparin
therapy is essential in maintaining hemostasis during certain procedures such
as bypass surgery; its administration poses significant risk to the patient.
Heparin may vary twelve-fold in both patient sensitivity and the rate at which
it disappears from the blood.
Overdosing heparin can result in dangerous bleeding, while under dosing
can lead to thrombosis.
The
Hemochron (Models 401 and 801) is a portable, battery operated instrument
designed to perform whole blood coagulation tests. The Model 401 instrument has one test well for single or
sequential test determinations. The
Model 801 instrument has two individual test wells allowing for simultaneous
performance of two coagulation tests.
The operation of both the 401 and the 801 is the same.
The
ACT is a test in which fresh whole blood is added to a test tube containing an
activator (celite glass particles or kaolin) and timed for the formation of a
clot. The type of activator used may
affect the degree of prolongation of the ACT.
The
Hemochron clot detection mechanism consists of a precision aligned magnet
within a test tube and a magnetic detector located within the test well. When a
test tube is inserted into the well, the magnetic detector senses a magnet
within the test tube as the tube slowly rotates. When the clot begins to form, it causes the magnet to lift within
the tube. Since the magnet has been
displaced, it is no longer sensed by the instrument's magnetic detector. The
instrument gives an audible beep and displays the coagulation time.
III. Materials:Hemochron Model 401 and 801
Instrumentation:
Location |
Model |
Serial # |
CMH # |
Test Tube
Type |
Daily QC |
ECMO |
801 |
T12460-8 |
1 |
P214/P215 |
Electronic |
ECMO |
401 |
T11607-4 |
2 |
P214/P215 |
Electronic |
Dialysis |
401 |
T12546-4 |
3 |
P214/P215 |
Electronic |
Dialysis |
401 |
T12618-4 |
4 |
P214/P215 |
Electronic |
CVS |
401 |
T12620-4 |
5 |
CA510 |
Electronic |
CVS |
401 |
T12622-4 |
6 |
CA510 |
Electronic |
PICU |
401 |
R5233-4 |
7 |
P214/P215 |
Electronic |
Cath Lab |
801 |
R5138-8 |
8 |
CA510 |
Whole Blood |
Syringes
for sample collection:
Tuberculin
syringe with no anticoagulant
One
of the following test tubes:
§
CA510 (black flip-top) - store at room temperature - No
expiration date.
§
P214/P215 (clear plastic flip-top) - store at room
temperature - No expiration date.
IV. Specimen Collection and Handling:
Prior
to collecting the patient’s sample, the patient’s identity must be verified by
looking at the arm band. Obtain blood
specimen. Use a 5 cc syringe to collect fresh whole blood. Adhere to the
appropriate technique below:
A.(CA510):
Indwelling venous blood line:
NOTE: Do not obtain blood from
heparinized access line, lock or indwelling heparin lock.
§
Discontinue fluids drip, if required.
§
Use a two-syringe technique -perform a 5 cc draw and discard
the first syringe. Obtain a 3 cc sample with the second syringe for testing.
Extracorporeal blood line port:
§
Flush the extracorporeal blood access line by withdrawing 5
cc of blood then obtain a 3 cc sample with a second syringe for testing.
Venipuncture:
§
Use a two-syringe technique and discard the first 2 cc
draw. Obtain a 3 cc sample with the
second syringe for testing.
Vacuum Draw:
§
Perform venipuncture maintaining the arm in its downward position.
§
Push tube to the end of the holder and pierce the top of the
tube. The contents of the tube should
not come in contact with the needle or stopper to prevent the possibility of
backflow from the tube.
§
Remove the tourniquet once blood begins to flow into the
tube.
§
The appropriate blood volume has entered the tube when the
blood flow into the tube ceases.
B.(P214/P215):
Draw 0.4 ml blood sample from a
previously flushed extracorporeal line using a 1 cc tuberculin syringe. Make certain that the blood sample is
representative (do not take sample from a line through which heparin has been
administered).
Extracorporeal blood line port:
§
Flush the extracorporeal blood access line by withdrawing 3
cc of blood.
V. ACT
Procedure:
A.(CA510)
1. From the collection syringe dispense
exactly 2.0 ml of blood into the test tube.
At the same time, depress the "start" button of the Hemochron
test well as blood first appears in the tube.
2. Close the flip-top. Immediately agitate
the test tube vigorously from end-to-end ten times to disperse the activator.
3. Insert test tube into the Hemochron test
well and rotate the tube clockwise until green "detector" light is
illuminated. Turn one additional revolution to assure the green light remains
on.
4. At the sound of the buzzer, record the
test result.
B.(P214/P215):
1. Before drawing the blood sample, hold
test tube vertically and tap the tube bottom several times on a horizontal
surface to shake activator particles down into testing zone.
2. Flip open the cap of the Hemochron
plastic test tube and dispense 0.4 ml of blood into the test tube. At the same time, depress the
"start" button as blood first appears in the test tube.
3. Close the flip-top. Holding the test tube in an upright position
with thumb and forefinger, mix blood and activator by gently flicking the
bottom of the tube with remaining fingers.
4. Insert test tube into the Hemochron and
rotate the tube clockwise until green "detector" light is
illuminated. Turn one additional revolution to assure that the green light
remains on.
5. At the sound of the buzzer, record the
test result.
VI. Quality Control:
The
quality control products are intended to verify proper instrument and test
performance, as well as operator technique.
Materials
Needed:
§
Hemochron Whole Blood Liquid Controls:
(Cat. No. CPL-2: Cardiac Cath, CVS)
(Cat. No. CPL 1: ECMO, Dialysis, PICU)
§
Hemochron Instrument 801 or 401
§
Plastic syringe (5 cc) with 20 or 21 gauge needle
§
Hemochron Coagulation test tubes
§
Electronic QC
A. Whole Blood Quality Control:
1. Handling and Storage:
Whole Blood controls are stable until
the marked expiration date when stored at 4-8°C.
Room temperature storage is possible for non-punctured vials; room
temperature dating is to a maximum of seven weeks, but must never exceed the
marked expiration date.
Note: If kept refrigerated, allow all quality
control and coagulation test vials and/or test tubes to first reach room
temperature and then stabilize for at least 20 minutes before use. This
may take up to 60 minutes.
2. Frequency:
If Electronic QC’s are not used, two
levels of quality control (normal and
abnormal) must be performed every 8 hour shift that patient testing
occurs. Controls should also be run
after any maintenance procedures and whenever the patient values do not match
the expected results. It is the
responsibility of the personnel performing the patient testing to perform
quality control.
If Electronic QC’s are used, two levels
of whole blood QC are used to validate each box of ACT test tubes. Once a box is validated for use, print or
affix a “VERIFIED” label on the outside of the box.
3. Whole Blood Quality Control Test
Procedure:
a. Peel back the flip top on the distilled,
deionized water vial, but do not remove the crimp seal (use this technique when
opening each vial). Using a 5 cc plastic syringe with needle, withdraw exactly
2 cc from the distilled, deionized vial and transfer into the plasma vial by
direct puncture of the stopper. Start
the Hemochron timer by depressing the start button. Remove the filled syringe from the plasma vial.
b. Using moderate end-to-end inversion,
agitate the plasma vial until 30 seconds elapses on the Hemochron display
screen to achieve rehydration. Avoid excessive agitation which will cause
foaming of the plasma. Verify complete
rehydration of the control visually.
c. Restart the Hemochron timer and allow the
plasma to stabilize for 60 seconds.
d. While the plasma is stabilizing, use the
same syringe and needle to withdraw exactly 1 cc from the calcium chloride vial.
e. After 60 seconds transfer 1 cc of calcium
chloride into the plasma vial by direct puncture of the stopper. Do not remove the syringe and needle from
the vial.
f. Firmly holding the syringe in place, vigorously agitate the vial
approximately
four to six times.
g. Using the syringe already in place,
withdraw the correct amount of the mixture for transfer into the appropriate
Hemochron test tube:
CA510- 2 cc
P214/P215- 0.5 cc
h. For
CA510 test tubes: Transfer the 2 cc into the Hemochron test tube. Depress the timer button. Vigorously agitate the tube four to six
times. Place the tube in the test well,
and then gently rotate the tube clockwise until the green "detector"
light is illuminated. Rotate the tube
one additional turn and assure that the green light remains on.
For
P214/P215 test tubes: Transfer 0.5 cc into the Hemochron test tube. Depress the
start button. Holding the tube vertically, flick the tube 4 to 6 times. Place the tube in the test well and then
gently rotate the tube clockwise until the green "detector" light is
illuminated. Rotate the tube one additional revolution to assure the green
light remains on.
I. When the test is complete, read the ACT
and record the value on the appropriate Quality Control Log Sheet. There are two types of quality control
logsheets for whole blood controls.
4. Troubleshooting:
If quality control does not fall within
the acceptable limits the following steps should be taken:
a. Verify that the quality control materials
and test tubes are within expiration date.
Check tubes for cracks and adequate activator. Verify that the controls and reagents are stored correctly.
Verify operator technique. Repeat test
using fresh control material.
b. If test is still out-of-limits, perform
weekly maintenance and repeat control.
c. If control remains out-of-limits, DO NOT USE INSTRUMENT, put a sign on
instrument stating that it is out of order. Contact Pathology department for
further assistance (Point of Care Medical Technologist ext. 4431 or pager
#4431). Use a backup instrument until
problem is resolved (see page 2 for list of instruments and locations).
d. Document all corrective actions on the
quality control log.
B. Electronic Quality Control:
1. Handling and Storage:
Electronic QC Verification Tubes are
stored at room temperature. They are
reusable and have no expiration date.
2. Frequency:
Two levels (normal and abnormal) are
used every eight hour shift of patient testing. The normal level is a 100 second QC tube (Cat# EQC-N). There are two abnormal levels. Those areas using CA510 tubes will use the
500 second abnormal QC tube (Cat# EQC-AII) and those using P214/P215 tubes will
use the 300 second QC tube (Cat# EQC-AI).
3. Electronic Quality Control Test Procedure:
a. Choose the appropriate QC verification
tubes.
b. Insert the electronic QC tube into the
test well.
c. Simultaneously press the START buttons on
the electronic QC tube and the Hemochron test well.
d. When the test is complete, read the ACT
and record the value on the Electronic
Quality Control and Maintenance log sheet.
Acceptable values are ± 10 seconds.
4. Troubleshooting:
If quality control does not fall within
the acceptable limits the following steps should be taken:
a. Repeat
Electronic QC Tube.
b. If a fault message is illuminated next to
the test result, the battery in the verification tube is nearly depleted and
should be replaced.
The verification tubes require 2 “AA”
batteries.
c. If control remains out-of-limits, DO NOT USE INSTRUMENT, put a sign on
instrument stating that it is out of order. Contact Pathology department for
further assistance (Point of Care Medical Technologist ext. 4431 or pager
#4431). Use a backup instrument until
problem is resolved (see page 2 for list of instruments and locations).
d. Document all corrective actions on the
quality control log.
VII. Maintenance:
Document all maintenance performed on
monthly maintenance sheet.
A. Weekly:
1. Inspect and clean the test well and test
tube drive collar of residual dried blood or other foreign matter. Use a cotton swab moistened with a 10%
bleach solution. Remove all moisture
with dry cotton swab after cleaning.
2. Verify that heater light is lit.
B. Monthly:
Check function of incubator. This can be performed by any operator or a
Medical Technologist. The temperature of the test wells is verified using the Hemochron Temperature Calibration
Test Tube. Place tube in well and
press "Start". If the instrument has not been in use, press the
"Prewarm" button prior to
temperature check. The temperature of each well should be 37± 0.5 °C. If temperature is not within limits, contact
the Pathology department. Use backup instrument (see page 2 for list
of instruments and locations.) Document temperature results on the maintenance
log.
C. Six Month:
1. Check function of timer against a
stopwatch.
2. Check clot detector sensitivity
(Calibration Verification).
Note: the two Model 801 test wells are
calibrated independently. Once calibration check is completed on the first or
upper well, simply insert a test tube into the second well, rotate clockwise
two complete revolutions, then repeat steps 3 through 6 for the second well.
a. Place the instrument on a flat level
surface.
b. Depress the "Prewarm"
switch 13 times in rapid succession.
the instrument detector sounder will "beep" and FAULT 04
briefly appears on the display screen followed by C-1. The instrument is ready to begin calibrating.
CAUTION: Now is the
only time to abort the calibration mode if desired. The entire calibration procedure must be followed once step 3 is
performed. To abort the calibration
mode, depress the "Prewarm" switch 13 additional times in rapid
succession. the instrument will
"beep" three times and the timer will begin. Depress the "Start" button to exit
the calibration mode.
c. To begin calibration, insert a test
tube into the appropriate test well.
Rotate the test tube two complete revolutions. Depress the
"Start" button. The "-" (hyphen) will now flash. This flashing
is common to each of the four calibration steps.
NOTE: If a
procedural error is made during calibration, the calibration sequence can be
restarted by depressing the "Prewarm" button (channel 1 on Model 801)
while the hyphen is flashing.
Once this step is complete, the instrument
will "beep" twice and C-2
will appear on the display screen.
d. With the tube still in place, prop
up the right side of the instrument at
approximately 45° angle. Once the instrument is securely in this
position, depress the "Start" button and the second calibration step
will begin. At completion, the
instrument will "beep" twice and C-3
will appear on the display screen.
e. With the test tube still in place,
stand the Hemochron against a flat surface (ie, a wall) so that the instrument
is at a 90N angle (perpendicular to the surface). Once the instrument is in position, depress the "Start"
button and the third calibration step begins.
The instrument will "beep" and C-4 will appear on the display screen.
f. Return the instrument to its correct
position. Remove the test tube. Depress
the "Start" button to begin the final calibration step. The instrument will "beep" four
times upon completion of calibration. A “0” will flash on the display
screen. The instrument is ready to
perform a coagulation test.
C. Other:
Instrument malfunctions that cannot be
corrected by routine maintenance procedures should be referred to the
Biomedical Engineering Department or the manufacturer for service.
International Technidyne Corporation 1-800-631-5945.
VIII. Quality Assurance:
Quality control results and maintenance
is reviewed weekly by a Medical Technologist.
Quality assurance reports are completed and distributed monthly to each
testing area for review.
IX. Limitations:
The
Hemochron ACT tests are affected by poor technique, including specimen
collection and test procedure. Inadequate specimen/reagent mixing may result in
reduced precision and accuracy.
Improper storage of test kits may affect results.
The
ACT may be affected by hemodilution, cardioplegic solutions, hypothermia,
platelet dysfunction, hypofibrinogenemia, other coagulopathies, and certain
medications.
Test
results of 600 or more seconds imply a slow rate of fibrin formation. Results greater than 600 seconds should be
interpreted qualitatively only. Results of 80 seconds or less generally
indicate improper instrument starting technique.
X. Expected Values:
The
reference range for ACT is to be established by each patient care area.
Generally, a baseline is taken before heparin is given. The ACT is then expected to increase after
the dose is given.
This
increase depends upon how large the dose is and the duration of the
procedure. All ACT test results are put
into the patient’s chart and/or appropriate computer system: Sunquest, CareView,
or Siemens Cath-Cor. Critical values
are up to the assessment of the physician.
XI. Interpretation of Test Results:
Hemochron
test results are quantitative indices of the in vitro rate of coagulability of
the blood specimen tested. The results may be interpreted by the clinician as
an indication of the status of anticoagulant therapy at the time of sample
collection.
XII. Linearity:
The
linearity of the ACT test is 80-600 seconds.
Results out of the reportable range should be reported as <80 or >600
sec.
XIII. Proficiency Testing:
Proficiency
testing will be provided by the College of American Pathologists (CAP). Two specimens will be received three times a
year for analysis of activated clotting time.
A
passing grade is required to maintain proficiency. Only Hemochron operators may perform PT.
Results
will be received and reviewed by the Director of Clinical Pathology. Results will be communicated to each
department by the Point-of-Care Medical Technologist via the monthly Quality
Assurance reports.
XIV. Competency Checks:
Competency
checks are performed on each operator during initial inservice and yearly. Observance of incompetency will necessitate
re-inservicing and inability to use analyzer until competency is acceptable.
XV. References:
Hemochron
Models 401 & 801 Operator's Manual.
International Technidyne Corporation. February, 1991.
Hemochron
System Instruction and Technical Guide. September, 1983.
Activated
Clotting Time by Hemochron CA510/CA510, K-ACT/FTK-ACT, P214/P215, S412, Package
Insert. May, 1992.
Hemochron
Operating Instructions CA510 Test Tubes/P214/P215 Test Tubes. International
Technidyne Corporation. March, 1988.
International
Technidyne Corporation, 23 Nevsky Street, Edison, New Jersey 08820.