Procedure |
Activated
Clotting Time of Whole Blood on the Hemochron Response |
Prepared by |
Date Adopted |
Supersedes Procedure |
|
|
New
Procedure |
Review Date |
Revision Date |
Signature |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Distributed to |
No.
of Copies |
Distributed to |
No.
of Copies |
Pediatric
Intensive Care |
1 |
|
|
Pathology
Department |
1 |
|
|
|
|
|
|
|
|
|
|
PRECAUTION:
WHILE PERFORMING THIS PROCEDURE, THE FOLLOWING SAFETY
MEASURES MUST BE TAKEN: GLOVES MUST BE WORN; SMOKING, EATING, DRINKING,
APPLICATION OF COSMETICS, AND MANIPULATION OF CONTACT LENSES ARE PROHIBITED
IN ALL TECHNICAL WORK AREAS. REFER TO
THE UNIVERSAL PRECAUTIONS POLICIES. |
Complexity:
Moderate
I. Method:
The
activated clotting time (ACT) is measured with the Hemochron Response instruments.
The ACT test is commonly used for heparin anticoagulation monitoring during
bypass surgery, neonatal Extracorporeal Membrane Oxygenation (ECMO),
hemodialysis, radio frequency ablation, and critical care.
II. Principle:
Heparin
therapy is essential in maintaining hemostasis during certain procedures such
as bypass surgery; its administration poses significant risk to the patient.
Heparin may vary twelve-fold in both patient sensitivity and the rate at which
it disappears from the blood.
Overdosing heparin can result in dangerous bleeding, while under dosing
can lead to thrombosis.
The
ACT is a test in which fresh whole blood is added to a test tube containing an
activator (celite glass particles or kaolin) and timed for the formation of a
clot. The particular clotting activator
that is used influences the time required for clot formation. Celite is the standard ACT reAgent used for
high level heparin monitoring because of its excellent activating properties. However, serine protease inhibitors such as
aprotinin that may be administered to certain patients to decrease
postoperative bleeding can prolong the celite activated ACT. When aprotinin is on-board, a
kaolin-activated ACT tube should be used.
The
Hemochron Response is a portable, battery operated instrument designed to
perform whole blood coagulation tests.
The Hemochron clot detection mechanism consists of a precision aligned
magnet within a test tube and a magnetic detector located within the test well.
When a test tube is inserted into the well, the magnetic detector senses a
magnet within the test tube as the tube slowly rotates. When the clot begins to form, it causes the
magnet to lift within the tube. Since
the magnet has been displaced, it is no longer sensed by the instrument's magnetic
detector. The instrument gives an audible beep and displays the coagulation
time.
III. Materials:
Syringes
for sample collection:
Tuberculin
syringe with no anticoagulant
One
of the following test tubes:
§
CA510 (black flip-top) - store at room temperature - No
expiration date.
§
P214/P215 (clear plastic flip-top) - store at room
temperature - No expiration date.
IV. Specimen Collection and Handling:
Prior
to collecting the patient’s sample, the patient’s identity must be verified by
looking at the arm band. Obtain blood
specimen. Adhere to the appropriate
technique below:
A.(CA510):
Indwelling venous blood line:
NOTE: Do not obtain blood from
heparinized access line, lock or indwelling heparin lock.
§
Discontinue fluids drip, if required.
§
Use a two-syringe technique -perform a 5 cc draw and discard
the first syringe. Obtain a 3 cc sample with the second syringe for testing.
Extracorporeal blood line port:
§
Flush the extracorporeal blood access line by withdrawing 5
cc of blood then obtain a 3 cc sample with a second syringe for testing.
Venipuncture:
§
Use a two-syringe technique and discard the first 2 cc
draw. Obtain a 3 cc sample with the
second syringe for testing.
Vacuum Draw:
§
Perform venipuncture maintaining the arm in its downward
position.
§
Push tube to the end of the holder and pierce the top of the
tube. The contents of the tube should
not come in contact with the needle or stopper to prevent the possibility of
backflow from the tube.
§
Remove the tourniquet once blood begins to flow into the tube.
§
The appropriate blood volume has entered the tube when the
blood flow into the tube ceases.
B.(P214/P215):
Draw 0.4 ml blood sample from a
previously flushed extracorporeal line using a 1 cc tuberculin syringe. Make certain that the blood sample is representative
(do not take sample from a line through which heparin has been administered).
Extracorporeal blood line port:
§
Flush the extracorporeal blood access line by withdrawing 3
cc of blood.
V. ACT
Procedure:
1. Press
MENU 1 or MENU 2 (depending on the well to be used for the test) once. The first page of the main menu is
displayed.
2. Press 2.
The QC Selections menu is displayed.
3. Press 4.
A prompt will confirm a patient test is being run:
“Patient
Test”
4. Dispense
the sample into the test tube (see directions below for specific test tube
being used)
5. Insert
tube into test tube well
6. Press
START 1 or START 2. A beep will signal
the start of the test and timing of the test begins.
7. Enter
Operator ID when prompted
8. Enter
Patient ID when prompted
(CA510)
1. From the collection syringe dispense
exactly 2.0 ml of blood into the test tube.
At the same time, press START 1 or START 2 as blood first appears in the
tube.
2. Close the flip-top. Immediately agitate
the test tube vigorously from end-to-end ten times to disperse the activator.
3. Insert test tube into the Hemochron test
well. The tube is rotated automatically
in the well until clot formation is detected.
4. When clot formation is detected, the
instrument beeps three times and the test name, clotting time (in seconds) is
displayed. The results of the test will
be displayed until CANCEL, START 1, or START 2 is pressed.
(P214/P215):
1. Before drawing the blood sample, hold
test tube vertically and tap the tube bottom several times on a horizontal
surface to shake activator particles down into testing zone.
2. Flip open the cap of the Hemochron
plastic test tube and dispense 0.4 ml of blood into the test tube. At the same time, press START 1 or START 2
as blood first appears in the tube.
3. Close the flip-top. Holding the test tube in an upright position
with thumb and forefinger, mix blood and activator by gently flicking the
bottom of the tube with remaining fingers.
4. Insert test tube into the Hemochron test
well. The tube is rotated automatically in the well until clot formation is
detected.
5. When clot formation is detected, the
instrument beeps three times and the test name, clotting time (in seconds) is
displayed. The results of the test will
be displayed until CANCEL, START 1, or START 2 is pressed.
VI. Quality
Control:
The
quality control products are intended to verify proper instrument and test
performance, as well as operator technique.
Materials
Needed:
§
Hemochron Whole Blood Liquid Controls:
§
Hemochron Instrument
§
Plastic syringe (5 cc) with 20 or 21 gauge needle
§
Hemochron Coagulation test tubes
§
Electronic System Verification Tube
A. Whole Blood Quality Control:
1. Handling and Storage:
Whole Blood controls are stable until
the marked expiration date when stored at 4-8°C.
Room temperature storage is possible for non-punctured vials; room
temperature dating is to a maximum of seven weeks, but must never exceed the
marked expiration date.
Note: If kept refrigerated, allow all quality
control and coagulation test vials and/or test tubes to first reach room
temperature and then stabilize for at least 20 minutes before use. This
may take up to 60 minutes.
2. Frequency:
If Electronic QC’s are not used, two
levels of quality control (normal and abnormal) must be performed every 8 hour
shift that patient testing occurs.
Controls should also be run after any maintenance procedures and
whenever the patient values do not match the expected results. It is the responsibility of the personnel
performing the patient testing to perform quality control.
If Electronic QC’s are used, two levels
of whole blood QC are used to validate each box of ACT test tubes. Once a box is validated for use, print or
affix a “VERIFIED” label on the outside of the box.
3. Whole Blood Quality Control Test
Procedure:
a. Press
MENU 1 or MENU 2 (depending on the well to be used for the test) once. The first page of the main menu is
displayed.
b. Press
2. The QC Selections menu is displayed.
c. Press
1 or 2, corresponding to whether a normal or abnormal control is being
run. The QC Ranges menu for the
selected control level is displayed
d. Press
YES if correct. (If the current limit
is not correct, press 1, 2, or 3 depending on what needs to be changed. The current limit or lot number is
displayed. A flashing cursor is
positioned where the first character of the new lower limit or lot # is to be
entered. If needed, enter the new value
and press YES.)
e. Dispense
the control sample into the test tube as described below under “Dispensing of
liquid QC into the tube.”
f. Press
START 1 or START 2. A beep will signal
the start of the test, and timing of the test begins.
g. Mix
the contents of the control test tube and place the tube in the test tube
well. The tube is rotated until clot
formation is detected.
h. If
the results pass, the instrument will display:
“Acceptable”
“Remove Tube”
And routine testing can
continue.
i. If
the results are unacceptable, the problem must be resolved and controls run
successfully before patient testing can be performed.
Dispensing of
Liquid QC into the tube:
a. Peel back the flip top on the distilled,
deionized water vial, but do not remove the crimp seal (use this technique when
opening each vial). Using a 5 cc plastic syringe with needle, withdraw exactly
2 cc from the distilled, deionized vial and transfer into the plasma vial by
direct puncture of the stopper.
b. Using moderate end-to-end inversion,
agitate the plasma vial for 30 seconds to achieve rehydration. Avoid excessive
agitation which will cause foaming of the plasma. Verify complete rehydration of the control visually.
c. Allow the plasma to stabilize for 60
seconds.
d. While the plasma is stabilizing, use the
same syringe and needle to withdraw exactly 1 cc from the calcium chloride
vial.
e. After 60 seconds, transfer 1 cc of
calcium chloride into the plasma vial by direct puncture of the stopper. Do not remove the syringe and needle from
the vial.
f. Firmly holding the syringe in place, vigorously agitate the vial
approximately
four to six times.
g. Using the syringe already in place,
withdraw the correct amount of the mixture for transfer into the appropriate
Hemochron test tube:
CA510- 2 cc
P214/P215- 0.5 cc
h. For
CA510 test tubes: Transfer the 2 cc into the Hemochron test tube.
Vigorously agitate the tube four to six times.
For
P214/P215 test tubes: Transfer 0.5 cc into the Hemochron test tube. Holding the
tube vertically, flick the tube 4 to 6 times.
4. Troubleshooting:
If quality control does not fall within
the acceptable limits the following steps should be taken:
a. Verify that the quality control materials
and test tubes are within expiration date.
Check tubes for cracks and adequate activator. Verify that the controls and reagents are stored correctly.
Verify operator technique. Repeat test
using fresh control material.
b. If test is still out-of-limits, perform
weekly maintenance and repeat control.
c. If control remains out-of-limits, DO NOT USE INSTRUMENT, put a sign on
instrument stating that it is out of order. Contact Pathology department for
further assistance (Point of Care Medical Technologist ext. 4431 or pager
#4431). Use a backup instrument until
problem is resolved (see page 2 for list of instruments and locations).
d. Document all corrective actions on the
quality control log.
B. Electronic Quality Control:
1. Handling and Storage:
The Electronic System Verification Tube
is stored at room temperature. It is
reusable and has no expiration date.
2. Frequency:
Two levels (normal and abnormal) are used
every eight hour shift of patient testing.
The normal level is a 100 second QC tube. There are two abnormal levels.
Those areas using CA510 tubes will use the 500 second abnormal QC tube
and those using P214/P215 tubes will use the 300 second QC tube.
3. Electronic Quality Control Test
Procedure:
a. Press
START 1 or START 2 (depending on the well to be verified) to begin a test on a
well. A beep will signal the start of
the test. At the same time the START
button is pressed, press the 100 second button on the Electronic System
Verification tube.
b. Insert
the electronic QC tube into the test well.
c. Enter the four digit Operator ID and
press YES
d. Enter the serial number of the ESV tube
(The serial number is located on the back of the ESV tube.)
e. At the completion of the test, the result
is automatically printed
f. Repeat the test, using the 300 or 500
second button for the first well. Then,
repeat the entire test on the second well.
The test
results are acceptable if all of the results are within 10 seconds of
the selected
time.
NOTE: If the
barcode label of the ESV tube is not read, display the QC Selections menu by
pressing MENU 1 or MENU 2 (depending on the well to be used for the test) once
to display the first page of the main menu and then pressing 2. Select 3 and enter the ESV serial number.
4. Troubleshooting:
If quality control does not fall within
the acceptable limits the following steps should be taken:
a. Repeat ESV Tube.
b. If a fault message is illuminated next to
the test result, the battery in the verification tube is nearly depleted and
should be replaced.
The verification tubes require 2 “AA”
batteries.
c. If control remains out-of-limits, DO NOT USE INSTRUMENT, put a sign on
instrument stating that it is out of order. Contact Pathology department for
further assistance (Point of Care Medical Technologist ext. 4431 or pager
#4431). Use a backup instrument until
problem is resolved (see page 2 for list of instruments and locations).
VII. Maintenance:
Document all maintenance performed on
monthly maintenance sheet.
A. Weekly:
1. Inspect and clean the test well and test
tube drive collar of residual dried blood or other foreign matter. Use a cotton swab moistened with a 10%
bleach solution. Remove all moisture
with dry cotton swab after cleaning.
2. Verify that heater light is lit.
B. Every 6 months:
Check function of incubator. This can be performed by any operator or a
Medical Technologist. The temperature of the test wells is verified using the
Hemochron Temperature Calibration Test Tube.
a. Press MENU 1 or MENU 2 (depending on the
well to be tested) and then press 3.
The Prewarm Menu is displayed.
b. Press 5.
Prewarming commences and the elapsed prewarming time of t 500 seconds is
displayed.
c. Insert the Temperature Verification Tube
into the test well.
d. When 0 seconds is displayed, read the
thermometer in the well.
e. The
temperature of each well should be 37± 1.0 °C. If temperature is not within limits, contact
the Pathology department.
f. Repeat the test for the second well.
Check
function of timer against a stopwatch.
VIII. Quality
Assurance:
Quality control results and maintenance
is reviewed weekly by a Medical Technologist.
Quality assurance reports are completed and distributed monthly to each
testing area for review.
IX. Limitations:
The
Hemochron ACT tests are affected by poor technique, including specimen
collection and test procedure. Inadequate specimen/reagent mixing may result in
reduced precision and accuracy.
Improper storage of test kits may affect results.
The
ACT may be affected by hemodilution, cardioplegic solutions, hypothermia,
platelet dysfunction, hypofibrinogenemia, other coagulopathies, and certain
medications.
Test
results of 600 or more seconds imply a slow rate of fibrin formation. Results greater than 600 seconds should be
interpreted qualitatively only. Results of 80 seconds or less generally
indicate improper instrument starting technique.
A
test is terminated and operation of the test well automatically shuts off if
clot formation is not detected within 1500 seconds after the test is
started. If the instrument terminates a
test in this manner, a FAULT-1500 message is displayed indicating the test
result is outside of the specified range.
X. Expected Values:
The
reference range for ACT is to be established by each patient care area.
Generally, a baseline is taken before heparin is given. The ACT is then expected to increase after
the dose is given
This
increase depends upon how large the dose is and the duration of the
procedure. All ACT test results are put
into the patient’s chart and/or appropriate computer system: Sunquest,
CareView, or Siemens Cath-Cor. Critical
values are up to the assessment of the physician.
XI. Interpretation of Test Results:
Hemochron
test results are quantitative indices of the in vitro rate of coagulability of
the blood specimen tested. The results may be interpreted by the clinician as
an indication of the status of anticoagulant therapy at the time of sample
collection.
XII. Linearity:
The
linearity of the ACT test is 80-600 seconds.
Results out of the reportable range should be reported as <80 or
>600 sec.
XIII. Proficiency
Testing:
Proficiency
testing will be provided by the College of American Pathologists (CAP). Two specimens will be received three times a
year for analysis of activated clotting time.
A
passing grade is required to maintain proficiency. Only Hemochron operators may perform PT.
Results
will be received and reviewed by the Director of Clinical Pathology. Results will be communicated to each
department by the Point-of-Care Medical Technologist via the monthly Quality
Assurance reports.
XIV. Competency Checks:
Competency
checks are performed on each operator during initial inservice, after 6 months
of employment, and yearly thereafter.
Observance of incompetency will necessitate re-inservicing and inability
to use analyzer until competency is acceptable.
XV. References:
Hemochron
Models 401 & 801 Operator's Manual. International Technidyne Corporation. February, 1991.
Hemochron
System Instruction and Technical Guide. September, 1983.
Activated
Clotting Time by Hemochron CA510/CA510, K-ACT/FTK-ACT, P214/P215, S412, Package
Insert. May, 1992.
Hemochron
Operating Instructions CA510 Test Tubes/P214/P215 Test Tubes. International
Technidyne Corporation. March, 1988.
International
Technidyne Corporation, 23 Nevsky Street, Edison, New Jersey 08820.