Procedure |
URINE
SPECIFIC GRAVITY BY REFRACTOMETER |
Prepared by |
Date Adopted |
Supersedes Procedure |
|
|
|
Review Date |
Revision Date |
Signature |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Distributed to |
No.
of Copies |
Distributed to |
No.
of Copies |
Pathology
Department |
1 |
Nursing
Units |
4 |
|
|
|
|
|
|
|
|
|
|
|
|
PRECAUTION:
WHILE PERFORMING THIS PROCEDURE, THE FOLLOWING SAFETY
MEASURES MUST BE TAKEN: GLOVES MUST BE WORN; SMOKING, EATING, DRINKING,
APPLICATION OF COSMETICS, AND MANIPULATION OF CONTACT LENSES ARE PROHIBITED
IN ALL TECHNICAL WORK AREAS. REFER TO
THE UNIVERSAL PRECAUTIONS POLICIES. |
CLIA
Complexity: Moderate
I. Principle
The
refractive index of urine correlates closely with the specific gravity. It is a
ratio of the velocity of light in air to the velocity of light in solution, and
is directly proportional to the number of dissolved solids in urine. This ratio varies directly with the number
of dissolved particles in urine and, as such, varies similarly with the
specific gravity of urine.
II. Equipment
Refractometer
Disposable pipet
III. Specimen
Follow
policy for random urine collection.
Prior to collecting the patient’s sample, the patient’s identity must be
verified by looking at the arm band.
Collect urine in a clean container and test it as soon as possible. Do not centrifuge or use preservative. Refrigerate urine if testing is delayed.
IV. Procedure
1. Place a drop of urine in the designated
spot on top of the refractometer.
2. Hold the instrument toward a source of
light and read the value directly from the specific gravity scale.
3. Clean the refractometer after each use
with distilled water or weak bleach
solution (Quanto), wipe lens with lens paper or soft gauze.
V. Results
Report
result to the nearest thousandth, (e.g., 1.010.)
Results
greater than 1.035 should be reported as > 1.035.
VI. Quality
Control
1. Reagents
§
Sterile water
§
Quantimetrix Normal Urine control - Supplied liquid,
ready-to-use, requiring no reconstitution.
It is prepared from human urine fortified to target levels with
compounds that produce the desired reaction when tested. Preservatives including sodium azide have
been added to inhibit microbial growth.
§
MAS Level 1
Liquid Urinalysis Control Solution
2. Storage of Quantimetrix Urine Control
a. The Liquid controls are stable until
expiration date stated on label when stored between 2°and 8°C. Controls
can be stored at room temperature for up to one month; record date opened and
expiration date on dropper vials on initial use. DO NOT FREEZE.
b. Discard the control if turbid or any
evidence of microbial contamination is present. Reddish brown uric acid crystals may form in the urine matrix and
precipitate due to changes in the temperature.
These crystals will not affect the performance of the control material.
3. Daily Quality Control
Once a day, a normal and abnormal urine
control should be tested on the refractometer.
a. On initial use, remove the controls from
the refrigerator and allow to reach room temperature prior to testing.
b. Remove cap and invert bottle. Apply urine control to refractometer as you
would a patient sample.
c. Read the refractometer according to the
procedure above. Record result on the Quality Control Log Sheet.
d. Wipe off dropper tips and recap control. The controls can be stored at room
temperature. Record room temperature
expiration date in the space provided on the bottle label.
Once a month,
the refractometer is checked for accuracy by measuring the specific gravity of
water (considered a standard) which should read 1.000 ±0.001. To verify the accuracy at a high specific
gravity, the MAS Level 1 Liquid
Urinalysis Control Solution is run.
4. Expected Range
Compare the lot number given on the
quality control log with the lot number on the bottle of urine control
material; they should match. The
expected results for the lot are designated on the log sheet for each testing
parameter.
If the readings with water consistently
depart from zero more than 0.001, the zero setting may need adjustment. Consult operative manual for instructions.
5. Troubleshooting
a. Clean
refractometer and repeat control.
b. If control still out-of-range, check
control for expiration date, inspect instrument, review technique.
c. Repeat control.
d. If control still out-of-range, open a new
bottle of control solution and repeat test.
e. If control comes in range, discard
controls that gave out-of-range results.
If controls are still out-of-range, go to step f.
f. Do not use instrument, call or page the
Point-of-Care Testing Coordinator at ext. 4431 or pgr. 4431.
VII. Proficiency Testing
As
an additional assurance of quality, two proficiency testing samples will be
distributed three times annually. These
samples will be from the Clinical Microscopy Survey (CM) from CAP and will be
distributed by the Pathology Department.
Only one select nursing unit will perform the survey per PT event. Results of the PT will be reported on the Urinalysis
Quality Management Report.
VIII. Reference Range
Specific
gravity = 1.003-1.030
The
refractometer gives values for specific gravity comparable to those obtained by
other methods with the advantage of using only one drop of urine. Results greater than 1.035 should be
reported as > 1.035. Critical Values
are up to the assessment of the physician.
IX. Limitations
IVP
dyes interfere with specific gravity resulting in abnormally high
readings. If the reading goes beyond
the scale, dilute the urine 1:2 and remeasure.
The presence of protein and glucose in urine also increases the specific
gravity.
X. Maintenance
1. Daily
a. Clean the eyepiece, prism glass and
coverplate with lens paper or soft gauze.
Any smears on any of these parts interfere with readings. Do not use any abrasive material or scratchy
paper to clean these parts.
2. Quarterly
Remove the rubber holder and wash it
out to make sure urine does not collect inside and corrode the refractometer.
3. Other
a. The meter is compensated to give correct
readings directly on aqueous solutions at all temperatures ranging from 15°C to 37°C. Never expose instrument to temperatures
above 65°C.
b. The instrument is damaged by immersion of
the eyepiece and focusing ring in water.
c. An air bubble to accommodate thermal
expansion of liquid is kept in a bubble trap at the end of the refractometer
cavity. If it escapes and appears in
the visible part of the prism, hold the instrument vertically, eyepiece down,
and shake lightly.
XI. Competency Checks
Competency
checks are performed at initial orientation, after 6 months of employment,
and yearly thereafter by the POC
Medical Technologist or a validated nurse.
XII. References:
1. Freeman JA, Beeler MF: Laboratory
Medicine: Clinical Microscopy. Lea and Febiger, Philadelphia, 1974:181-189.
2. Ames Company: Modern Urinalysis.
Ames Company, 1973:19-22.
3. Davidson I, Henry JB (eds.): Clinical
Diagnosis by Laboratory Methods, 15th ed. WB Saunders, Philadelphia,
1974:45-47.
4. Lippman RW: Urine and Urinary
Sediment. Charles C. Thomas, Springfield, IL. 1957:101-103.
5. The dropper Plus, Package Insert,
Quantimetrix Corporation, Hawthorne, CA. October, 1993.