Procedure |
GLUCOSE
AND KETONES IN URINE WITH CHEMSTRIP UGK URINE TEST STRIPS |
Prepared by |
Date Adopted |
Supersedes Procedure |
|
|
New
Procedure |
Review Date |
Revision Date |
Signature |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Distributed to |
No.
of Copies |
Distributed to |
No.
of Copies |
Nursing
Department |
1 |
|
|
Pathology
Department |
1 |
|
|
|
|
|
|
|
|
|
|
PRECAUTION:
WHILE PERFORMING THIS PROCEDURE, THE FOLLOWING SAFETY
MEASURES MUST BE TAKEN: GLOVES MUST BE WORN; SMOKING, EATING, DRINKING,
APPLICATION OF COSMETICS, AND MANIPULATION OF CONTACT LENSES ARE PROHIBITED
IN ALL TECHNICAL WORK AREAS. REFER TO
THE UNIVERSAL PRECAUTIONS POLICIES. |
CLIA Complexity: Waived-Definitive
I. Principle
Chemstrip
uGK test strips are for the semiquantitative visual estimation of glucose and
ketones in urine.
Glucose:
The glucose reagent area has two
distinct zones of different reagent concentrations specifically designed to
provide optimal sensitivity to urine glucose levels ranging from 100 to 5,000
mg/dL. Both zones of the reagent area
contain the enzymes glucose oxidase and peroxidase, and the color indicator
tetramethylbenzidine. D-glucose is
oxidized to gamma-D-gluconolactone with glucose oxidase acting as a
catalyst. The hydrogen peroxide that
results from the reaction oxidizes the indicators in the presence of
peroxidase. The intensities of the
colors formed are proportional to the glucose concentration of the specimen.
Ketones:
The detection of ketone bodies in urine
is based on the method attributed to Legal.
II. Specimen
Prior
to collecting the patient’s sample, the patient’s identity must be verified by
looking at the arm band. If it is an outpatient, verbally verify patient’s name
and date of birth with patient or parent/guardian. Collect urine in a clean container that will allow complete
immersion of test strip area. Test as soon as possible after collection. Do not centrifuge specimen. If testing cannot be performed within one
hour after collection, the specimen should be refrigerated immediately at 2-80C
and returned to room temperature prior to testing. Mix thoroughly before testing.
III. Reagents
Chemstrip
uGK Urine Test Strips (Cat. No. 00513).
Store at room temperature below 30°C; do not
freeze.
Label vial with date opened; do not use
after expiration date.
Caution:
Test strips contain nitroferricyanide.
Do not touch the reagent pads with skin or mucous membranes; if contact
occurs flush with large amounts of water.
Timer
or watch with a second hand
IV. Procedure
1. Collect fresh urine specimen in clean,
dry container. Mix well before testing.
2. Remove one test strip from bottle and
replace cap immediately. Completely
immerse reagent area in urine and remove immediately to avoid dissolving out
reagents.
3. While removing, run the edge of the strip
against the rim of the urine container to remove excess urine.
4. Start timing for two minutes. After two
minutes, match the colors on the test strip with the color blocks printed on
the vial label. Any color changes after
three minutes are not important.
5. DO NOT COMPARE TEST STRIP FROM ONE
VIAL WITH THE COLOR CHART ON ANOTHER
VIAL (DIFFERENT LOT NUMBER). THIS COULD CAUSE MISINTERPRETATION. EACH COLOR CHART IS MADE TO MATCH THE TEST
STRIPS FOR THAT LOT NUMBER.
6. When the colors of the glucose test patch
are between the color blocks on the vial, then glucose test result is between
those color blocks. Compare both
glucose color blocks at the same time.
7. If
the ketone turns from beige to violet, the test is positive.
V. Quality Control
A. Reagents
Quantimetrix urine dipstick controls,
normal and abnormal level; prepared from human urine to which is added
predetermined amounts of chemicals. Liquid, ready-to-use.
B. Storage
and Stability
Stable until expiration date stated on
label, when stored between 2° to 8°C. Controls can be stored up to one month at
room temperature. Label control vials
with open date and room temperature expiration date. DO NOT FREEZE.
C. Procedure
1. On initial use, remove the control tube
from the refrigerator and allow to reach room temperature prior to testing.
2. Remove cap and invert bottle. While holding reagent strip, gently squeeze
the sides of the dropper bottle and touch the tip of the bottle to the test
pad. Thoroughly saturate the reagent
pads. Turn the dipstick on its side and
drain excess control onto absorbent material.
3. Read
the dipstick according to the procedure above.
4. Wipe off dropper tips and recap
controls. The controls can be stored at
room temperature for up to one month when not in use.
5. Record
results on the Quality Control Log Sheet.
D. Frequency
Two levels of quality control (normal
and abnormal) should be run once a week.
Controls should also be run whenever a new vial of uGK Reagent Strips is
opened.
E. Troubleshooting
If quality control results are not
within the acceptable limits the following steps should be taken:
1. Confirm that the strips and controls are
within the opened expiration dates shown on the label.
2. Repeat test with fresh control material.
If still out of range, go to step 3.
3. Repeat
test with new vial of test strips.
4. Discard
any controls or reagent strips which gave invalid test results.
5. Do not use strips for patient testing if
quality control results are out of range.
VI. Expected Results
Glucose
or ketones should not be detected in normal urine. A positive reaction indicates a need for further diagnostic
evaluation of the patient. Critical
Values are up to the assessment of the physician.
VII. Limitations
Glucose:
False negative or falsely lowered
glucose readings may occur rarely at high concentrations of ascorbic acid. Highly colored urine (i.e. due to certain
medications or bilirubin) may the color reaction and prevent accurate
readings. Glucose values must be
determined within three minutes of starting the test.
Ketones:
2-Mercaptoethane sulphonate sodium and
other sulfhydryl-containing compounds may cause false positive results. Ketone values must be determined within the
specified three minute time limits.
VIII. Competency Checks
Competency
checks are performed during at initial orientation, after 6 months of
employment, and yearly thereafter by the POC Medical Technologist or a
validated nurse.
IX. References
1. Legal
E. Chem Centr. 14:652 (1883)
2. Zweig
M. Clin. Chem. 32/4, 674-677 (1986)
3. Chemstrip uGK Urine Test Strips, package
insert. Boehringer Mannheim
Diagnostics. Indianapolis, IN. 1988.
4. The dropper Plus Urine Dipstick Control
Kit, package insert. Quantimetrix Corporation, Hawthorne, CA. September, 1993.