|
|||
Hosted by Medical Automation Systems |
|||
|
REFERENCES |
CAP
Regulations
CAP checklists were updated in September/October. Here are the new links: Point of Care Testing
Blood Gas
Limited Service Laboratory
Lab General
CAP Information New
Checklists Includes Many Changes The new POCT checklist includes many new changes. In addition, CAP has clarified a lot of question in their commentary and defined a lot of new terms. If interested in comparing the new and old version of the checklists, visit www.cap.org/html/ftpdirectory/compare.html.According to CAP, Point of Care Testing refers to those analytical patient tests provided within the institution, but performed outside the physical facilities of the clinical laboratories. The central criterion of POCT is that it does not require permanent dedicated space. This type of testing is performed by a large number of different non-laboratory healthcare personnel who have other patient care responsibilities in addition to POCT. Another key aspect that distinguishes POCT from main laboratory testing is that it utilizes instruments and kits that are hand-carried or otherwise transported to the vicinity of the patient for immediate testing at that site. POCT does not include limited service satellite laboratories with fixed dedicated testing space; these are covered under the Limited Service Laboratory Checklist. When
records are maintained centrally by a designated coordinator or
POCT Director, only one copy of this Point of Care Testing
Checklist need be completed.
The Inspector will review all centrally maintained records
and visit at least a sampling of the testing sites in order to
evaluate compliance with the standards.
If records are not maintained centrally, the Inspector must
visit each POCT site, and a separate checklist must be completed
for each location. In
the latter case, each POCT site will be billed as an additional
laboratory section. To
be accredited, all analytes being measured under the POCT
program/site must be included in the on-site inspection and comply
with the CAP Standards for Laboratory Accreditation.
POCT programs may be inspected as sections of the central
laboratory if they are registered under the same CLIA-88 number.
In this circumstance, they are included in the Laboratory
General Checklist used for the central laboratory.
If the POCT sites are registered under separate CLIA
numbers, a separate Laboratory General Checklist must be completed
for each POCT program.
Quality
Control
Quality
control samples are samples that act as surrogates for patient
specimens. They are
to be periodically processed like a patient sample to monitor the
ongoing performance of the entire analytic process, and must be
judged acceptable prior to reporting patient results.
They must be evaluated and organized such that they can be
used daily by the testing personnel or supervisory technical staff
to detect problems, trends, etc.
It is implicit that quality control specimens be tested in
the same manner as patient specimens.
Moreover, QC specimens must be analyzed by personnel who
routinely perform patient testing-this does not imply that each
operator must perform QC daily, so long as each instrument/test
system has QC performed at required frequencies. Most
quantitative tests are traditionally monitored with two levels of
liquid QC. This is to
be done at a frequency within which the accuracy and precision of
the measuring system is expected to be stable, but at least each
day that patient testing is performed.
The daily use of two levels of liquid QC may not be
required for certain test systems where the daily use of
electronic controls is demonstrably sufficient.
For
hematology, coagulation, and blood gas tests, two different levels
of control material are required during each eight hour period of
patient testing and each time there is a change of reagents.
For
qualitative tests, a positive and negative control must be
included with each run of patient specimens. As an exception to this general practice, the specific
frequency of such testing for multiparameter urine chemistry
dipsticks may vary according to workload and testing location, and
may not occur with each run.
However, the frequency must be defined and followed by the
laboratory. For
direct bacterial or viral antigen tests that include internal
positive and negative controls, a positive and negative external
control must be tested with each new kit lot number or separate
shipments of a given lot number.
Internal quality control tests can be used for subsequent
testing, based upon the manufacturers’ recommendations.
Oversight review of quality control must occur at least monthly by the Laboratory Director or designee. Beyond the specific CAP requirements, a laboratory may perform more frequent review at intervals that it determines appropriate for its setting and the assays involved. Because of the many variables, CAP makes no specific recommendations on the frequency of any additional assessment/review of QC. Proficiency
Testing
POCT programs
must participate in a CAP-approved proficiency program of
interlaboratory comparison testing appropriate to the scope of
their program. Proficiency testing is required for all analytes regardless
of their CLIA classification.
When possible, the proficiency testing samples must be
processed as patient samples, and analyzed by personnel who
routinely test patient samples, using the same method.
Further, these samples should be periodically rotated among
all analyzers used for testing, and among all testing personnel.
Rea
|
|
[
Home | Search
| Index | Feedback
]
Last updated: 02/22/2002
Questions or corrections: Webmaster
© 2000 Medical
Automation Systems, Inc. Legal Notice.